Dakota Ultrasonics
74
25) Press the
ENTER
key to display the measurement screen.
26) Press the
key to automatically adjust the settings and measure the
calibration standard.
27) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll the hot menu items until
ACCEPT CAL
is highlighted.
28) Press the
ENTER
key to accept the calibration and return to the
One Point
Zero
list, or
ESC
to cancel the calibration.
29) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
One Point Zero
list until
CALC ZERO
is
highlighted.
30) Press the
ENTER
key to calculate, change, and display the new velocity
and zero values, or
ESC
to cancel the calibration.
Note:
If
ENTER
was pressed to calculate the new values, pressing the
ESC
key once will display the
CAL
menu items to review the new
ZERO
value
calculated.
31) Finally, press the
MEAS
key to return to the measurement screen and
begin taking readings.
Typically when the zero value is off and the hardware has not changed, it’s because
the temperature of the bolt has changed. In this case, the zero value should not be
changed, but the temperature value should be. Therefore the bolt becomes sort of a
temperature sensor in itself. This is also a good way to monitor temperature changes
in the bolt. The procedures below demonstrate how to adjust the temperature to
match the zero value.
Adjusting the Temperature to Match the Zero Value