Dakota Ultrasonics
22
3.6 Measuring Reference Lengths
At this point, the
MAX II
is setup and ready to start measuring reference lengths.
We’ve already setup a bolt group to store the reference length data, and now need to
display the group storage locations prior to making measurements.
Storing a Reading
1) Press the
MEAS
key once to activate measure menu items. Press the
MEAS
key multiple times to move down and the
ESC
key multiple times to
move up until the
LOG
cell is highlighted. Alternatively, press the
UP
and
DOWN
arrow keys multiple times using the left button wheel.
Note:
Pressing the
key will automatically open the
Group View Box
.
2) Press the
ENTER
key to display the
Group View Box
.
3) Press the
UP, DOWN, LEFT, & RIGHT
arrow keys using the left button
wheel to scroll the target cell cursor to the desired storage location.
Note:
The approximate lengths must always be located in the first column A for
all bolts in the group.
4) Place a small amount of couplant on the transducer and attached it to the
bolt. Make sure that the transducer is seated so that there is no couplant
Summary of Contents for MAX II
Page 2: ......
Page 7: ......
Page 8: ......
Page 54: ...Dakota Ultrasonics 46 Figure 6 Reflection in a bending bolt ...
Page 215: ......