Dakota Ultrasonics
90
Verifying the Probe Zero at a Later Time
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right, and the
ESC
key multiple times to tab left,
until the
DATA
menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT
and
RIGHT
arrow keys multiple times
using the left button wheel.
2) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the sub menu items until
OPEN
is highlighted.
3) Press the
ENTER
key to display the
Open Group List Box.
4) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the list menu items until the
group name
containing the probe zero data
CAL DATA
is highlighted.
5) Press the
ENTER
key to load the zero data group into memory, followed by
pressing the
OK
key to confirm loading the group into memory.
Note:
The user will always use this group for the zero verification procedure,
as it contains the original settings and measurement data of the probe zero
calibration standards. Multiple data groups may have been created for different
transducer types and frequencies.
6) Apply a drop of couplant to the cal standard or transducer, and attach it to
one end of the cal standard. Rotate the transducer clockwise and counter