WARNINGS 

Read Before Operating

 

The MultiRAE User’s Guide must be carefully read by all 

individuals who have or will have the responsibility of 

using, maintaining, or servicing this product. The product 

will perform as designed only if it is used, maintained, and 

serviced in accordance with the manufacturer’s instructions. 

 

CAUTION! 

Never operate the monitor when the rear cover is re-

moved. Remove rear cover, sensors, and/or battery only 

in an area known to be non-hazardous. Never use the 

instrument with the calibration adapter installed, as this 

can cause distorted readings, a potential safety threat. 

 

Note: 

If the MultiRAE is equipped with a gamma 

sensor, it comes pre-calibrated from the factory and no 

calibration is required. You can challenge it anytime 

with a radioactive check source. 
 

User Interface 

 

The MultiRAE’s user interface consists of the display 

and three keys, [Y/+], [MODE] and [N/-]. The flippable 

LCD displays information such as monitored threats, 

real-time readings, measurement units, alarm type (when 

in alarm, including cal. overdue), battery and pump 

status, datalog (if on), and radio and connection quality 

(if available). 

 

 

 

 

 

Charging The MultiRAE 

 

Always fully charge the battery before use. Contacts on 

the bottom of the MultiRAE meet the Travel Charger’s 

or Charging Cradle’s contact pins, transferring power. 

Make sure the charger and  MultiRAE are firmly 

attached. Then connect the AC Adapter’s plug to the 

charger, and plug its transformer into an AC outlet. 

While charging, the LED on the cradle glows red. 

When the battery is fully charged, the LED glows 

green. 

 

 

 

Turning The MultiRAE On 

 

With the instrument turned off, press and hold the 

[MODE] key until the audible alarm stops, and then 

release. During startup, the battery, buzzer, vibration 

alarm, and LEDs are tested, and then the MultiRAE 

performs self-testing of its other functions. When the 

main measurement screen appears, the MultiRAE is 

ready for calibration or use. 

 
Note: 

If the battery is completely empty, then the 

display briefly shows the message “Battery Fully 

Discharged,” and the MultiRAE shuts off. You should 

charge the battery or replace it with a fully charged 

battery before turning it on again. 

 

Note:

 If Fast Startup is enabled on the instrument, 

fewer screens are shown during startup, compared 

to Normal Startup sequence. 

 

Turning The MultiRAE Off 

 

Press and hold [MODE]. A 5-second countdown to shut-

off begins. You must continue pressing on the key for the 

entire shutoff process. If you remove your finger from the 

key during the countdown, the shutoff operation is 

canceled and the MultiRAE continues normal operation. 

 

When the countdown ends and the screen displays “Unit 

Off,” release your finger from the [MODE] key. The 

MultiRAE is now off. 

 

 

Testing The Alarms 

Under normal-operation mode and non-alarm condi-

tions, the buzzer, vibration alarm, LED, and backlight 

can be tested anytime by pressing [Y/+] once. If any 

alarm does not respond, check the Alarm Settings in 

Programming Mode to make sure all alarms are 

enabled. If any alarms are enabled but are not 

functional, do not use the instrument. 

 

2. 

3. 

5. 

6. 

4. 

Charging 
status LED 

[N/-] key 

[MODE] key 

Gas inlet 

Display 

LEDs 

Alarm 
Buzzer 

LEDs 

[Y/+] key 

Charging and 
Communication 
Contacts 

External filter 

AC adapter 
plug 

Cable from 
AC adapter 

MultiRAE 

Charger 

Phone: 

408-952-8200      

Fax: 

408-952-8480 

Email:

 [email protected] 

www.raesystems.com 

RAE Systems by Honeywell 
3775 N. First St. 
San Jose, CA 95134-1708 USA 

Reviews: