background image

Bump Testing and Calibration Setup 

 

The MultiRAE can be automatically bump tested and 

calibrated using the AutoRAE 2 Test and Calibration 

System (refer to its User’s Guide for instructions). 

Manually calibrate using a fixed-flow regulator (flow rate 

between 0.5 and 1.0 liters per minute) and the supplied 

special calibration adapter that covers the gas inlet: 

 

1.

 

Connect the gas cylinder, flow regulator, tubing 

(must use Teflon tubing for PID sensor), and 

calibration adapter to the MultiRAE.  

2.

 

With the MultiRAE in Normal Mode, enter 

Programming Mode by pressing and holding  both 

[MODE] and [N/-] until the password screen appears.  

3.

 

Input the 4-digit password.  (The default password is 

“0000.”  If you do not know the password, select 

“Done.”) Then follow the menus to select single- or 

multi-sensor bump test, zero, or span calibration. 

Important! 

After a bump test or calibration, remove the 

calibration adapter to ensure correct readings. 

Zero & Fresh Air Calibration 

 

The MultiRAE should be zero-calibrated in clean air 

with 20.9% oxygen or with a cylinder of clean zero air. 

In Programming Mode, select “Fresh Air.” Then: 

 

Press [Y/+] to start a Fresh Air calibration for the listed 

sensors. All are fresh-air calibrated at once. 

 

To individually zero calibrate sensors: 

1.

 

Select “Single Sensor Zero” and select a sensor. 

2.

 

Press [Y/+] to select a sensor to zero calibrate. 

3.

 

Start the flow of the zero gas, if used, and press [Y/+]. 

4.

 

The screen says, “Zeroing” and counts down. 

5.

 

When done, it says, “Zero Calibration Passed” (the 

reading should be 0 or very close to it for VOC and toxic 

gas sensors, and 20.9% Vol. for an oxygen sensor). 

6.

 

Shut off the flow of zero air (if used) and remove 

the calibration adapter. 

Note: 

If your MultiRAE is equipped with a CO

2

 sensor, 

it must be zero calibrated using 100% Nitrogen (N

2

), or 

isobutylene, instead of fresh air or zero air. 

Span Calibration 

 

In Programming Mode, and with “Multi Sensor Span” or  

“Single Sensor Span” highlighted: 

 

1.

 

Press [Y/+]. The screen displays the sensor(s) to be 

calibrated.  

          · Multi: The list is shown. 

          · Single: Select a sensor and press [Y/+] 

2.

 

Attach the calibration adapter, and connect the 

calibration gas cylinder’s flow regulator to the 

MultiRAE, and start the gas flow. 

3.

 

Press [Y/+] to start calibration. 

4.

 

Upon completion, a pass/fail calibration result 

appears and the readings are shown (they should be 

within ±10% of the span gas value).  

Note:

 If a VOC sensor is installed, a second 

calibration can be performed to enhance linearity, 

requiring different calibration gas. 

5. Turn off the gas and remove the calibration  

    adapter. 

 

Bump (Functional) Testing 

 

Bump test to confirm that the sensors and alarms are 

functional.  

 

Important!

 Test the alarms, as described in panel 6 

(above), prior to performing a bump test. 

 

Important! 

Make sure all sensors have warmed up 

before performing the bump test. You can tell a 

sensor has warmed up if you see a reading next to it 

name on the display. If it has not warmed up, you see 

three dashes next to it.  

 

With the MultiRAE in Normal Mode: 

 

1.

 

Enter the Bump Test menu. Follow instructions 

in panel 7 (above) or use the easy shortcut: Press 

both the [Y/+] and [N/-] buttons at the same time 

and hold them for 5 seconds. The Multi-Bump 

menu then appears.

 

 

2.

 

The bump test process consists of two steps, each 

requiring its own calibration gas. The LEL and O

2

 

sensors are tested first, followed by the PID sensor. 

Press [Y/+] to start the bump test. While the bump 

test is being performed, the readings for each sensor 

are shown.  

3.

 

Once the bump test completes, pass/fail results are 

shown for each sensor.  

4.

 

Press OK to proceed to the PID sensor test. 

Connect Isobutylene gas and press [Y/+] to start the 

test. After the test completes, pass/fail results are 

shown. 

5.

 

Press OK to return to the main measurement 

screen. 

6.

 

Turn off the gas flow. 

7.

 

Remove the calibration adapter. 

 

BATTERY PACKS 

 

Li-Ion battery pack (PN: M01-3051-000 or M01-

3053-000) and an alkaline battery adapter (PN: 

M01-3052-000 or M01-3054-000) are supplied 

with each MultiRAE. 

 

There are two types of output power for battery 

packs or adapters. The battery pack (PN: M01-

3051-000) and adapter (PN: M01-3052-000) are 

used for MultiRAE model number PGM-62x0. 

Battery pack (PN: M01-3053-000) and adapter 

(PN: M01-3054-000) are used for model number 

PGM-62x6/PGM-62x8. 

 

The alkaline battery adapter accepts four AA 

alkaline batteries (use only Duracell MN1500). Do 

not mix old and new batteries or batteries from 

different manufacturers. 

8. 

9. 

10. 

7. 

11. 

P/N: M01-4016-000 Rev C  201403 