MAX II
Bolt Tension Monitor
87
21) Press the
key to automatically setup and measure the 6” cal bar.
Note:
Once an echo is found and the
MAX II
is measuring the cal bar, the
GAIN
, as well as the
POLARITY
may need to be adjusted.
22) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll the hot menu items until
ACCEPT CAL
is highlighted.
23) Press the
ENTER
key to accept the calibration and return to the
Two Point
Zero
list, or
ESC
to cancel the calibration.
24) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
Two Point Zero
list until
CALC ZERO
is
highlighted.
25) Press the
ENTER
key to calculate the zero and velocity values.
26)
Make note of the velocity value calculated, as it will need to be changed in
the MATL menu prior to saving the measurements in the CAL DATA group.
27) Press the
ESC
key to review the
ZERO
value calculated.
28) Press the
MENU
key multiple times to tab right and the
ESC
key multiple
times to tab left until the
MATL
menu is highlighted and displaying the
submenu items. Alternatively, press the
LEFT
and
RIGHT
arrow keys
multiple times using the left button wheel.
29) Press the
UP
and
DOWN
arrow keys multiple times using the left
button wheel to scroll through the
sub menu items until
VELOCITY
is
highlighted.
30) Press the
ENTER
key to display the
Digits Edit Box
.
31) Press the
UP
and
DOWN
arrow keys multiple times using the left or
right button wheel to scroll the highlighted value.