MAX II
Bolt Tension Monitor
5
D.
Alarm
– Indicates that the alarm feature has been activated and currently in
use.
E.
RF A-Scan Display (split screen)
– Displays the radio frequency sound wave
reflection returned the opposite end of the bolt being measured. The RF view
displays both the positive and negative cycles.
F.
Quantity
– Displays the current quantity setting used (elongation, load, stress,
strain or time).
G.
Digital Measurement
– Display in inches, millimeters, KSI, KIPS, %strain,
time (nanoseconds).
H.
Rectified A-Scan Display
– Displays either the positive or negative half cycle
wave reflection returned from the opposite end of the bolt being measured,
depending on the polarity selected.
I.
Time & Date
– Feature to display either the time, date, or both time and date
on the main measurement screen.
J.
Hot Menu items
– We call this menu section our “hot menu”, as these items
are the most commonly adjusted features, requiring quick access from the
user. They can be displayed and scrolled by pressing the MEAS key at any
time. The MEAS key advances forward and the ESC key backwards to the
next hot menu item. Alternatively, the
UP
,
DOWN
,
LEFT
and
RIGHT
arrow keys on the left button wheel can be used to either scroll through the
sub menu items, or advance to the next HOT MENU items list. There are 4
HOT MENU lists that contain the most commonly adjusted items from the
tabbed menus, and in the same order as the tabbed menus. Note: Some of
the lists contain multiple tabbed menu items in order to condense the number
of HOT MENU lists.
K.
Units
– Displays the current measurement units being used (English, Metric).
L.
Gain & Step Size
– The gain and increment step size can be continuously
used and adjusted at any time using the
UP
and
DOWN
arrow keys on
the right button wheel to scroll the value, and the
GAIN
key to activate the
Digit Edit
screen to adjust the step size.
M.
Scan Bar
– Another view of bolt length in a deflection style horizontal bar.
This is a visual tool that would enable the user the ability to see the length
from a graphics perspective.
N.
Large Digital Measurement
– Display in inches, millimeters, KSI, KIPS,
%strain, time (nanoseconds).
O.
Feature Status Bar
– Indicates the features currently enabled and in use in
the following order:
Measurement
Mode
Alarm
Capture (screen shots)
Quantity
Temperature
Summary of Contents for MAX II
Page 2: ......
Page 7: ......
Page 8: ......
Page 54: ...Dakota Ultrasonics 46 Figure 6 Reflection in a bending bolt ...
Page 215: ......