The digitizer feature provides additional stability when to the measurement by
increasing the number of samples. This can be beneficial when measuring shorter
length fasteners with units set to in/mm hr (high resolution), in order to increase the
resolution. The reverse scenario also then applies to longer length fasteners, not
requiring the increased sample size/shots. The digitizer has three setting options of
1, 2, 4x, with 4x providing the highest number of samples/shots. It should be noted
that increasing the digitizer sample slows down the update speed. For shorter bolts,
this won’t be as noticeable as using a 4x digitizer setting for longer bolts. The
procedure to change the digitizer setting is outlined below:
For quick access and editing, refer to navigating the Hot Menus page 36.
Setting the Digitizer
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
TUNE
menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT
and
RIGHT
arrow keys multiple times
using the left button wheel.
2) Use the
UP
and
DOWN
arrow keys multiple times using the left button
wheel to scroll through the sub menu items until
DIGITIZER
is highlighted.
3) Use the
LEFT
and
RIGHT
arrow keys multiple times on the right button
wheel to scroll the
DIGITIZER
options.
4) Once the desired
DIGITIZER
setting is displayed, press the
MEAS
key to
return to the measurement screen.