7 Performing maintenance
266
bplogic intelligent pump controller
18/02/2021
7.4 Fault table
Error description
Cause
Remedy
System does not start.
No operating system on the
flash drive, no external drive
with operating system
Insert a USB storage
medium or a microSD
card containing the
current operating
system.
on
Power supply faulty
Check the power supply.
No WLAN function
WLAN off
on page 243
No WLAN dongle inserted
WLAN is not projected
on page 204
Metallic cover plate inhibits
radio waves
Use a plastic cover
plate.
Incorrect date and time
entries
Pump controller was de-
energised for an extended
period.
commissioning after an
extended shutdown
on
NTP server cannot be
reached.
on page 245
No write authorisation
Another user is already
logged in.
Log the other user out.