M470039-01 REV-F
Page 110
CR1
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CR1
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR1
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be displa y on the front pa nel LCD.
CR HIGH RANGE TRANSIENT LEVEL CALIBRATION
CR high ra nge transient level ca libra tion requires a power supply that ha s high volta ge a nd medium current, but opera tor
ha s to estima te the power consumption of loa d.
Use “
” and “
” until you see TRAN CR2 parameter. Press enter:
CR2
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR2
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CR2
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR2
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be displa y on the front pa nel LCD
.
CP RANGE TRANSIENT LEVEL CALIBRATION
CP high ra nge tra nsient level ca lib ra tion requires a power supply tha t ha s enough power output. For exa mple, a PLW6K-
60-150 model needs a t lea st 30V/90A power output to ca libra te the unit. Only qua lified personnel should a ttempt this
ca libra tion, due to the high level of power.
Use “
” and “
” until you see TRAN CP0 parameter. Then press enter:
CP0
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP0
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CP0
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP0
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be display on the f ront panel LCD.
Use “
” and “
” until you see TRAN CP1 parameter. Press enter:
CP1
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP1
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CP1
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP1
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be display on the f ront panel LCD.
Use “
” and “
” until you see TRAN CP2 pa rameter. Press enter:
CP2
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP2
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CP2
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP2
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be display on the f ront panel LCD.
Summary of Contents for Amrel PLA 2U
Page 2: ......
Page 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Page 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Page 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Page 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Page 141: ...M470039 01 REV F Page 140...
Page 161: ......
Page 162: ......