Built in pulse genera tor for continuous, pulsed, a nd toggled tra nsient opera tion.
•
Seven progra mma ble protection modes: Over Volta ge Protection (OVP), Under Volta ge Protection (UVP),
Over Current Protection (OCP), Under Current Protection (UC P), Overpowe r Protection (OPP), a nd Under
Power Protection (UPP).
•
Fa n speed control for reduced a coustic noise under light loa d conditions.
•
0-10Vdc a na log progra mma ble.
•
Simple closed-ca se ca libra tion.
•
Highest power density: up to 18KW in 2U height
•
Field-upgra dea ble ca pa bility for GPIB, RS232, USB a nd Ethernet.
•
Three ra nges for ea ch opera ting mode; Auto or Ma nua l
•
Externa l a na log progra mming control for ea ch mode a nd ra nge
•
Pa nel lock ca pa bility
•
Progra m the Slew rise time a nd Slew fa ll time independently
•
50us tra nsient time a t CC mode; 500us tra nsient time a t CV mode
•
Anti-Oscilla tion controls: System ba ndwidth controls
•
Oscilla tion detection a nd protection
•
Anti-Sa tura tion controls: Under Volta ge Lockout
•
16 bits progra mming a nd rea d ba ck ca pa bility
•
Cross mode a nd Cross ra nge ca pa bility; defa ult off
•
Remote Inhibit a nd Dry conta ct Fa ult output
•
Pulse power ca pa bility with Overpower Protection
•
Field-upgra dea ble firmwa re
•
Self-dia gnostic ca pa bility
•
Redunda nt Over Tempera ture Protection
•
Four Step profiles; 256 step points for ea ch profile
•
Soft-Sta rt Functiona lity
•
Convenient a nd efficient Hot Keys for mode a nd Tra nsient Mode Selection, Input a nd Tra nsient On/Off Control,
Slew Ra te Settings, Front pa nel trigger, a nd Short.
•
Independent La rge Numerica l Volta ge a nd Current LED displa y
GENERAL INFORMATION AND OPERATING NOTES
Operating Modes:
The four modes of opera tion a re:
Constant Current (CC):
The electronic loa d will sink a current in a ccorda nce with the progra mmed va lue rega rdless of the input volta ge.
Note 1:
Never use CC mode for testing a consta nt current source. The regula tion of the two units will fight for control of
the current a nd a n unsta ble oscilla tion will result.
Note 2:
During CC mode, the electronic loa d ca n opera te under the fa stest tra nsient time.
Constant Voltage (CV):
The electronic loa d will a ttempt to sink enough current to control the source volta ge to the progra mmed va lue. It a cts a s a
shunt volta ge regula tor when opera ting in the CV mode.
Note 1:
Never use CV mode for testing a consta nt volta ge source. The regula tors of the two devices will buck ea ch other
trying to ga in control of the volta ge which will lea d to a n unsta ble condition.
Note 2:
In
CV mode, the tra nsient time will be a djusted to a slower setting to ensure sta ble opera tion.
Summary of Contents for Amrel PLA 2U
Page 2: ......
Page 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Page 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Page 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Page 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Page 141: ...M470039 01 REV F Page 140...
Page 161: ......
Page 162: ......