M470039-01 REV-F
Page 113
SIX:
NEW ELOAD FEATURES
NEW ELOAD FEATURES
FW: 2.17/060216
•
Power Interruption (Current Interruption)
FW: 2.29/061005
•
Under-temperature Trip Caution Message (UTT)
•
Input Contactor Control
•
Ultra-low Range Option
FW: 2.40/061005
•
Step Modification
•
OSC Range Selection
FW: 2.49/081007
•
Interlock Function
•
PV Sweep Function
FW: 2.84/100608
•
Master-Slave Parallel Operation
FW: 2.90/130129
•
Step Time Unit Function.
•
New Overpower Function
1. POWER INTERRUPTION (CURRENT INTERRUPTION) MODE:
For Battery and Fuel Cell Testing, a common method used to measure the internal impedance is the
Current/Power Interrupt or just the Interrupt method. The battery or fuel cell is loaded at a constant current,
and the current is interrupted by bringing it down to zero or lower current level. During the interrupt period,
the battery or fuel cell voltage is measured. As a result, the magnitude of the impedance can be determined.
The Interrupt method can be achieved in either CC mode or CP mode, with recommendations in the CC mode.
Local Control:
Press SINGLE twice
➔
FLASHING
→
Set the eLoad in CC or CP Mode
→
Set the Main Level to 0Adc or the desired current level via the encoder
→
Set the Peak Level to a current level larger than the Main to load the DUT
Note: Peak Level is set via the PEAK hot key and the encoder for value edit
→
Set the Power Interruption Pulse Width (PINT: 200 μs ~ 25000 μs)
→
Set the System Bandwidth (OSC3, OSC2, OSC1 and DEFAULT)
→
Activate the INPUT
→
Activate the PULSE mode
→
Press TRIGGER to activate Power Interruption
Summary of Contents for Amrel PLA 2U
Page 2: ......
Page 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Page 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Page 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Page 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Page 141: ...M470039 01 REV F Page 140...
Page 161: ......
Page 162: ......