ONCE
Causes the STEP to advance only one point after each trigger. Triggers that arrive during a
dwell delay are ignored
■
AUTO
Causes the entire STEP to be executed sequentially after the starting trigger, paced by its
dwell delays. As each dwell delay elapses, the next point is immediately executed.
Command Syntax
STEP:CURRent:STATe <NR1>
Parameters
0 | OFF(0) | 1 | ON(1) | 2 | AUTO(2) | 3 | ONCE(3)
Query Syntax
STEP:CURRent:STATe?
Returned Parameters
<NR1>
0=OFF
1=ON
2=AUTO
3=ONCE
NOTE
: Plea se refer to the eLoa d Progra mming ma nual a nd Opera tion Ma nua l for further Step Opera t ion a nd
comprehensive Step SCPI Comma nds.
6. OSC Range Selection
The OSC Range selection now includes D DEFAULT, D OSC1, D OSC2 &
D OSC3. With the new selections, the eLoad provides more bandwidth selections for applications
with dc load + ac perturbation. The DISABLE OSC range will stop detection of oscillating conditions; in the
case ac + dc inputs are required for testing applications.
Product used with oscillation protection turned off may cause product failure. Failure occurring as a result of
oscillation application or turn-off of this feature will not be covered under warranty.
Local Control
MENU
LOAD CONTROL
➔
OSCILLATE
➔
OSC CONTROL
➔
DEFAULT
➔
OSC1
➔
OSC2
➔
OSC3
➔
DEFAULT + DISABLE
➔
OSC1 + DISABLE
➔
OSC2 + DISABLE
➔
OSC3 + DISABLE
CAUTION
Summary of Contents for Amrel PLA 2U
Page 2: ......
Page 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Page 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Page 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Page 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Page 141: ...M470039 01 REV F Page 140...
Page 161: ......
Page 162: ......