M470039-01 REV-F
Page 107
Use “
” and “
” until you see MAIN CC0 parameter. Press enter:
CC0
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CC0
HI = CNT:
Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be displa y on the front pa nel LCD.
Press “
” to select MAIN CC1. Then press enter.
CC1
LO=
CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER
CC1
HI = CNT:
Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be displa y on the front pa nel LCD.
Press “
” to select MAIN CC2. Then press enter.
CC2
LO=
CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER
CC2
HI = CNT:
Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be displa y on the front pa nel LCD.
CONSTANT RESISTANCE (CR0) LOW RANGE MAIN LEVEL CALIBRATION
CR low ra nge ma in level ca libra tion requires a power supply tha t ha s low volta ge a nd high current, but operator ha s to
estima te the power consumption of loa d.
Use “
” and “
” until you see MAIN CR0 parameter. Press enter:
CR0
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR0
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CR0
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR0
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be displa y on the front pa nel LCD.
CONSTANT RESISTANCE (CR1) MIDDLE MAIN LEVEL CALIBRATION
CR middle ra nge ma in level ca libra tion requires a power supply that ha s medium volta ge a nd medium current, but
opera tor ha s to estima te the power consumption of electronic loa d.
Use “
” and “
” until you see MAIN CR1 parameter. Press enter:
CR1
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR1
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CR1
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR1
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be displa y on the front pa nel LCD.
CONSTANT RESISTANCE (CR2) HIGH RANGE MAIN LEVEL CALIBRATION
CR high ra nge ma in leve l ca libra tion requires a power supply tha t ha s high volta ge a nd medium current, but opera tor ha s
to estima te the power consumption of loa d.
Summary of Contents for Amrel PLA 2U
Page 2: ......
Page 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Page 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Page 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Page 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Page 141: ...M470039 01 REV F Page 140...
Page 161: ......
Page 162: ......