M470039-01 REV-F
Page 109
should be displa y on the front pa nel LCD.
Press “
” to select TRAN CV2. Then press enter.
CV2
LO=
CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER
CV2
HI = CNT:
Input measured voltage from Volt.DMM then press ENTER key. A “SUCCESS” message
should be displa y on the front pa nel LCD.
CC LOW/MIDDLE/HIGH RANGE, TRANSIENT LEVEL CALIBRATION
CC tra nsient level ca libra tion requires a power supply tha t covers full ra nge of high ra nge input current of loa d. Estima te
the power consumption of the electronic loa d, a nd a djust the volta ge input to a sa fe limit for the loa d.
Use “
” and “
” until you see TRAN CC0 parameter. Press enter:
CC0
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CC0
HI = CNT:
Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be displa y on the front pa nel LCD.
Press “
” to select MAIN CC1. Then press enter.
CC1
LO=
CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER
CC1
HI = CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be displa y on the front pa nel LCD.
Press “
” to select MAIN CC2. Then press enter.
CC2
LO=
CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER
CC2
HI = CNT:
Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be displa y on the front pa nel LCD.
CR LOW RANGE TRANSIENT LEVEL CALIBRATION
CR low ra nge tra nsient level ca libra tion requires a power supply tha t ha s low volta ge a nd high current, but opera tor ha s to
estima te the power consumption of electronic loa d.
Use “
” and “
” until you see TRAN CR0 parameter. Press enter:
CR0
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR0
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CR0
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR0
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be displa y on the front pa nel LCD.
CR MIDDLE RANGE TRANSIENT LEVEL CALIBRATION
CR middle ra nge tra nsient level ca libra tion requires a power supply tha t ha s medium volta ge a nd medium current, but
opera tor ha s to estima te the power consumption of electronic loa d.
Use “
” and “
” until you see TRAN CR1 parameter. Press enter
CR1
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
Summary of Contents for Amrel PLA 2U
Page 2: ......
Page 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Page 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Page 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Page 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Page 141: ...M470039 01 REV F Page 140...
Page 161: ......
Page 162: ......