Set up the logic analyzer
1
Set up the Channel menu.
a
Press the Config key. In the pop up menu, select Scope Channel.
b
Select the Input field, then select C1.
c
Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d
Select the Coupling field, then select 1M
Ω
/ DC.
C A U T I O N
Set the Channel Coupling field to 1M
Ω
/ DC or damage to the equipment will result.
e
Move the cursor to the s/Div field, then use the RPG knob to dial in 500 ns.
2
Set up the Display menu.
a
Press the Display key.
b
Select the Mode field, then select Average.
c
Move the cursor to the Average # field. Type 8 on the front-panel keyboard, then press
Done.
d
Select the Grid field and set it to On.
e
In the Waveform selection menu, delete channel 2. If channel 1 is not inserted, insert
channel 1.
3
Set up the Trigger menu.
a
Press the Trigger key.
b
Select the Mode/Arm field, then select Immediate.
Testing Performance
To test the voltage measurement accuracy (oscilloscope)
3–64