To test the CAL OUTPUT ports
This test does not need to be performed every time an operational accuracy
calibration is done. Testing the oscilloscope CAL OUTPUT ports checks the following:
•
DC CAL OUTPUT voltage
•
AC CAL OUTPUT voltage
This test verifies that the CAL OUTPUT voltages are operating within limits, so that
they can provide accurate calibration for the oscilloscope operational accuracy
calibration and probe calibration.
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Digital Multimeter
0.1 mV resolution, better than 0.005% accuracy
3458A
Cable
BNC (m)(m) 48-inch
8120-1840
Adapter
BNC (f) to Dual Banana Plug
1251-2277
Set up the equipment
Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes if you have not already done so.
Calibrating and Adjusting
To test the CAL OUTPUT ports
4–6