Recommended Test Equipment (logic analyzer)
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Use
*
Pulse Generator
150 MHz, 3.0 ns pulse width,
< 600 ps rise time
8133A Option 003
P,T
Digitizing Oscilloscope
≥
6 GHz bandwidth, < 58 ps rise time
54750A mainframe with
54751A plugin module
P
Function Generator
Accuracy
≤
(5)(10
−
6
)
×
frequency,
DC offset voltage
±
6.3 V
3325B Option 002
P
Digital Multimeter
0.1 mV resolution, 0.005% accuracy
3458A
P
BNC-Banana Cable
11001-60001
P
BNC Tee
BNC (m)(f)(f)
1250-0781
P
Cable
BNC (m)(m) 48 inch
8120-1840
P,T
SMA Coax Cable (Qty 3)
18 GHz bandwidth
8120-4948
P
Adapter (Qty 4)
SMA(m)-BNC(f)
1250-1200
P, T
Adapter
SMA(f)-BNC(m)
1250-2015
P
Coupler (Qty 4)
BNC (m)(m)
1250-0216
P, T
20:1 Probes (Qty 2)
54006A
P
BNC Coax Cable
BNC(m)(m), >2 GHz bandwidth, >1 meter length 10503A
T
BNC Test Connector, 17x2
(Qty 1)
**
P
BNC Test Connector, 6x2
(Qty 4)
**
P,T
Digitizing Oscilloscope
> 100 MHz Bandwidth
54600B
T
BNC Shorting Cap
1250-0774
T
BNC-Banana Adapter
1251-2277
T
*A = Adjustment P = Performance Tests T = Troubleshooting
**Instructions for making these test connectors are in chapter 3, "Testing Performance."
General Information
Recommended Test Equipment (logic analyzer)
1–9
Summary of Contents for 1670G Series
Page 20: ...1 12...
Page 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Page 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Page 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Page 201: ...Theory of Operation The Oscilloscope Board 8 11...