![Agilent Technologies 1670G Series Service Manual Download Page 140](http://html1.mh-extra.com/html/agilent-technologies/1670g-series/1670g-series_service-manual_2867830140.webp)
7
To exit the ROM Test, select Done. Note that the status changes to PASSED or
FAILED.
8
Install a formatted disk that is not write protected into the flexible disk drive.
Connect an RS-232-C loopback connector onto the RS-232-C port. Run the
remaining System Tests in the same manner.
9
Select the Front Panel Test.
A screen duplicating the front-panel appears on the screen.
a
Press each key on the front panel. The corresponding key on the screen will change
from a light to a dark color.
b
Test the knob by turning it in both directions.
c
Note any failures, then press the Done key a second time to exit the Front Panel Test.
The test screen shows the Front Panel Test status changed to TESTED.
10
Select the Display Test.
A white grid pattern is displayed. These display screens can be used to adjust the display.
a
Select Continue and the screen changes to full bright.
b
Select Continue and the screen changes to half bright.
c
Select Continue and the test screen shows the Display Test status changed to TESTED.