![Agilent Technologies 1670G Series Service Manual Download Page 83](http://html1.mh-extra.com/html/agilent-technologies/1670g-series/1670g-series_service-manual_2867830083.webp)
To test the input resistance (oscilloscope)
Testing the input resistance verifies the performance of the following specification:
•
Input resistance
This test checks the input resistance at the 50
Ω
and 1 M
Ω
settings in the Coupling
field.
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Digital Multimeter
Measure resistance (4-wire) better than 0.25%
accuracy
3458A
Cables (2)
BNC (m)(m) 48-inch
8120-1840
Adapter
BNC Tee (m)(f)(f)
1250-0781
Adapters (2)
BNC (f) to Dual Banana Plug
1251-2277
Set up the equipment
1
Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes if you have not already done so.
2
Set up the multimeter to make a 4-wire resistance measurement.
Testing Performance
To test the input resistance (oscilloscope)
3–59