To test the power supply voltages
Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and
assemblies.
W A R N I N G
Hazardous voltages exist on the power supply and the LCD display. Only service-trained
personnel who are aware of the hazards involved, such as fire and electrical shock, should
perform this procedure.
1
Turn off the instrument, then remove the power cable. Remove the cover of the
instrument and the disk drive assembly (including the support plate).
2
Reconnect both the hard disk drive and flexible disk drive to the CPU board
3
Connect the power plug, then turn on the instrument.
4
Check for voltages on the acquisition board power input connectors according to
the following diagram.
5
Turn off the instrument. Allow time for the capacitors in the power supply to
discharge. Remove and replace the power supply if needed.
6
Reassemble the instrument.
Troubleshooting
To test the power supply voltages
5–20