Set up the logic analyzer
1
Set up the Channel menu.
a
Press the Config key.
b
At the pop up menu, select Scope Channel.
c
Select the Input field, then select C1.
d
Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
e
Move the cursor to the V/Div field, then use the RPG knob to dial in 20 mV.
f
Move the cursor to the Offset field. Set the Offset to 0 V by typing 0, then pressing the
Select key.
g
Select the Coupling field, then select 50
Ω
/ DC.
2
Set up the Trigger menu.
a
Press the Trigger key.
b
Select the Mode/Arm field, then select Immediate.
Testing Performance
To test the input resistance (oscilloscope)
3–60