Set up the logic analyzer
1
Set up the Calibration menu.
a
Press the Waveform key.
b
Press the Waveform key again. At the pop up, select Scope Calibration.
c
Select the Mode field, then select Service Cal.
d
Select the Procedure field, then select Cal BNC Out.
e
Select Signal Probe Comp, then select DC Level.
f
Select the DC volts field, and set it to 0 V.
Calibrating and Adjusting
To test the CAL OUTPUT ports
4–7