4
Enable the pulse generator channel 1 output (with the LED off).
5
Set up the Waveform menu.
a
Press Run to fill the acquisition memory so the markers can be configured.
b
Press the Waveform key.
c
Move the cursor to the sec/Div field, then use the knob to dial in 100 ns.
d
Select the Markers Off field, then select Pattern.
e
Select the Specify Patterns field. Select X entering 1 and O entering 1.
f
Select Done to exit the Specify Patterns menu.
g
Move the cursor to the X-pat field. Type 1, then press Done.
h
Move the cursor to the O-pat field. Type 16384, then press Done.
i
Select O-pat from Trigger field, and then select from X marker.
Testing Performance
To test the time interval accuracy (logic analyzer)
3–57