Acquire the data
1
Press the RUN key. The clicking of attenuator relays should be audible. Verify
resistance readings on the digital multimeter of 50
Ω
±
0.5
Ω
(49.5 to 50.5
Ω
). Record
the reading in the performance test record.
2
In the Channel menu change the Coupling field to 1M
Ω
/ DC. The clicking of
attenuator relays should be audible.
3
Press the RUN key. Verify resistance readings on the digital multimeter of 1 M
Ω
±
10
k
Ω
(0.990 to 1.010 M
Ω
). Record the reading in the performance test record.
4
In the Channel menu change the Coupling field to 50
Ω
/DC and V/Div to 200 mV/Div.
Repeat steps 1 through 3.
5
In the Channel menu change the Coupling field to 50
Ω
/DC and V/Div to 1 V/Div.
Repeat steps 1 through 3.
6
In the Channel menu change the Coupling field to 50
Ω
/DC and V/Div to 4 V/Div.
Repeat steps 1 through 3.
7
Connect the male end of the BNC tee to the channel 2 input of the oscilloscope
module.
8
Repeat from "Set up the logic analyzer" for channel 2, replacing channel 1 with
channel 2 where applicable.
See Also
If a reading is not within limits, then the attenuator for the out-of-bounds channel should be
replaced (see chapter 6).
Perform an operational accuracy calibration
At this point, an operational accuracy calibration should be performed. Complete the
following procedures in chapter 4, "Calibrating and Adjusting."
•
"To test the CAL OUTPUT ports"
•
"To calibrate the oscilloscope"
•
"Self Cal menu calibrations"
Testing Performance
Perform an operational accuracy calibration
3–62
Summary of Contents for 1670G Series
Page 20: ...1 12...
Page 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Page 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Page 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Page 201: ...Theory of Operation The Oscilloscope Board 8 11...