4
Set up the Format menu.
a
Press the Format key.
b
Move the cursor to the field showing the channel assignments for the pod under test.
Press the Clear Entry key until the pod channels are all assigned (all asterisks (*)).
Press the Done key.
c
Select Master Clock, then select a double edge for the clock of the pod under test.
Turn off the other clocks.
d
In the Master Clock menu, select Setup/Hold, then select 4.0/0.0 ns for the pod being
tested. Select Done. Select Done again to exit the Master Clock menu.
Troubleshooting
To test the logic analyzer probe cables
5–30