4
Configure the trigger according to the connected channels.
a.
Press the Trigger key.
b.
Select the field next to the pattern recognizer "a" under the label Bus1. Type the
following for your logic analyzer, then press Select.
Agilent 1670G – "AA"
Agilent 1672G – "AA"
Agilent 1671G – "2A"
Agilent 1673G – "A"
5
Enable the pulse generator channel 1, channel 2, and trigger ouputs (with the LEDs
off).
Testing Performance
To test the multiple-clock, multiple-edge, state acquisition (logic analyzer)
3–34