To test the hard disk drive voltages
Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and
assemblies.
W A R N I N G
This procedure is to be performed by service-trained personnel aware of the hazards
involved, such as fire and electrical shock.
Equipment Required
Equipment
Critical Specification
Recommended
Model/Part
Digitizing Oscilloscope
> 100 MHz Bandwidth
54600B
1
Turn off the instrument, then remove the power cable. Remove the instrument cover
and the disk drive assembly.
2
Remove the two locking pins from the top of the power supply. Slide the CPU board
out of the instrument enough to access J12 on the outer edge of the CPU board.
3
Reconnect the disk drive cables.
4
Connect the power cable, then turn on the instrument.
5
In the Sys PV menu select the Disk test. Select Run-Repetitive.
Troubleshooting
To test the hard disk drive voltages
5–26