![Agilent Technologies 1670G Series Service Manual Download Page 39](http://html1.mh-extra.com/html/agilent-technologies/1670g-series/1670g-series_service-manual_2867830039.webp)
Test the 0 V User threshold
1
Set up the Format menu.
a
Press the Format key.
b
Select the field to the right of Pod A1, then select User in the pop-up menu, then enter
0 V.
2
Test the high-to-low transition.
a
On the DC power source, enter -0.099 V.
b
On the logic analyzer, press Run. The display should show all channels at a logic "0".
3
Test the low-to-high transition
a
On the DC source, enter 0.099 V
b
On the logic analyzer, press Run. The display should show all channels at a logic "1".
4
Record a PASS/FAIL in the performance test record for Threshold Accuracy Pod 1 -
0 V.
Testing Performance
To test the threshold accuracy (logic analyzer)