Set up the logic analyzer
1
Set up the Calibration menu.
a
Select the Signal field, then select Probe Comp.
2
Set up the Channel menu.
a
Press the Chan key.
b
Select the Coupling field, then select 1M
Ω
/ DC.
c
Move the cursor to the Probe field, press Select, then use the RPG knob to dial in 1:1.
Verify the AC CAL OUTPUT port
1
Using the BNC cable, connect channel 1 of the oscilloscope to the AC CAL OUTPUT
connector.
2
In the Calibration menu select Autoscale.
3
Press the Meas key. Verify that the waveform is approximately 0.8 Vp_p at
approximately 1.0000 KHz. Record the reading in the performance test record.
Calibrating and Adjusting
To test the CAL OUTPUT ports
4–9