Test the next pod
1
Using the 17-by-2 test connector and probe tip assembly, connect the data and clock
channels of the next pod to the output of the function generator until all pods have
been tested.
2
Unassign the channels just tested.
a
Press the Format key.
b
Move the cursor over the channel assignment field for the pod tested. Press Clear
Entry until all channels show unassigned (all ’.’).
c
Move the cursor over the Data on Clocks channel assignment field. Press Clear Entry
until all channels show unassigned (all ’.’).
3
Start with "Set up the logic analyzer" on page
3−
10, substituting the next pod to be
tested for pod 1.
Testing Performance
To test the threshold accuracy (logic analyzer)
3–16