Connect the Agilent 1672G or Agilent 1673G Logic Analyzer to the Pulse Generator
Testing
Combination
Connect to
8133A
Channel 2 Output
Connect to
8133A Channel 2
Output
Connect to
8133A Channel 1
Output
1
Pod 1, channel 3
Pod 2, channel 3
Pod 3, channel 3
Pod 4, channel 3
Pod 1, channel 11
Pod 2, channel 11
Pod 3, channel 11*
Pod 4, channel 11*
J-clock
*Agilent 1672G only
3
Activate the data channels that are connected according to one of the previous
tables.
a
Press the Format key.
b
Select the field showing the channel assignments for one of the pods being tested,
then press the Clear entry key. Using the arrow keys, move the selector to the data
channels to be tested, then press the Select key. An asterisk means that a channel is
turned on. When all the correct channels of the pod are turned on, press the Done
key. Follow this step for the remaining pods.
Testing Performance
To test the single-clock, single-edge, state acquisition (logic analyzer)
3–21