Checking the direction of rotation
at initial pump start-up
Before each pump start-up, check that oil levels of ADP and Roots housings
are visible in the middle of the sight glass: do this check with the pump
stopped.
In order to prevent moisture from entering the pump before installation, it
has been pressurized before hand with nitrogen and sealed with blank-off
flanges.
Remove the guards blocking the intake and exhaust holes; these
components prevent foreign bodies from entering the pump during
transport and storage. It is dangerous to leave them on a pump in
operation.
Before using the monitoring, check that the mechanical and electrical
connections defined in the chapter B have been made.
Connect
the cover plugs
J1
J14
In local mode, the pump will run only
if the cover plugs (delivered with the
pump) are respectively connected on
J1 and J14 connectors (remote control
connector).
In remote mode, if the emergency stop
is not interfaced, it is necessary to
connect the cover plug J1.