G 01069 - Edition 07 - July 12
1/4
C 43
Use of the M4 monitoring system for
pumping operation
Neutral gas purge control
Operation setting
Pump start-up
To start the Roots and ADP simultaneously, set the
CMD.ROOTS
menu
on DISABLED.
To start or stop the Roots while the ADP is running, set the
CMD.
ROOTS
menu on ENABLED.
The Roots will start or stop by pressing the
ROOTS
key.
When the Roots is running, a message “Roots” is displayed on the
hand held remote module.
For safety reasons, the pump is set with
CMD PURGE
on DISABLED.
This selection does not allow the user to stop the gas purge while
pump is running.
When the purge gas is selected, the message N
2
is displayed on the
hand held remote module.
Note:
in some process, it can be necessary to stop the gas purge
during pumping, set the CMD PURGE on ENABLED and the S2
coding wheel to “1” position. This setting concerns also the LM
versions.
According to the selected operating mode, check the position of the
S1 switches on the electrical interface, and at the rear of the hand-
held remote module (see
C 10
).
Check the direction of rotation at the initial start-up (see
B 51
).
The performance and the operational safety of this product is guaranteed
provided that it is used in normal operating parameters defined in this
manual.
Study the preliminary precautions (see
B 00
).
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two