G 03219 - Edition 01 - December 06
7/7
D 500
The pump is running and then is showing
a warning or fault message
Incident
Cause
Troubleshooting
Fault
KM1 contactor not closed
y
Check
CB2
(see
D 10
).
y
Check that there is a strap between
pins 2 and 3 of emergency stop
terminal (see
B 100
).
Other problem
Call customer service center.
Fault
RT2 breaker switched off
y
Check that ADS option is selected.
(see
C 42
).
y
Check
RT2
setting (see
D 10
).
y
Switch
RT2
on.
Other problem
Call customer service center.
Fault
RT1 breaker switched off
y
Check
RT1
setting (see
D 10
).
y
Switch
RT1
on.
y
Check the exhaut line.
Other problem
Call customer service center.
Fault
KM2 contactor not closed
y
Check
CB2
(see
D 10
).
y
Check that there is a strap between
pins 2 and 3 of emergency stop
terminal (see
B 100
).
y
Check the emergency stop switch
(
J1
).
Other problem
Call customer service center.
D36 ADP CONTACTOR
D36 ADP CONTACTOR
D37 ROOTS BREAKER
D37 ROOTS BREAKER
D38 ADP BREAKER
D38 ADP BREAKER
D39 ROOTS CONTACTOR
D39 ROOTS CONTACTOR
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two