2/3
D 10
First level of maintenance
G 02273 - Edition 04 - September 14
The diagnosis and troubleshooting guide (from
D 30
) may lead to
check the electrical components of the monitoring cabinet described
below:
Location of
electrical components
(electrical cabinet)
I1
Main switch
RT1
ADP
protection
RT2
ROOTS
protection
CB1
RT3
Transformer
protection
Heating
protection
Electrotechnics
protection
CB2
Electronic
links
W1
ADP
power measurement
KM1
ADP
command
KM2
ROOTS
command
Electronics
protection
CB3
Breaker rating
Pump
Voltage
RT1
RT2(1)
RT3
CB1
ADP 122 / ADS 602
B.T.
200 à 230 V - 50/60 Hz
10 A
10 A
1 A
1 A
H.T.
380 à 480 V - 50/60 Hz
6 A
6 A
0,7 A
0,7 A
ADS 1202
B.T.
200 à 230 V - 50/60 Hz
10 A
18 A
1 A
1 A
H.T.
380 à 480 V - 50/60 Hz
6 A
13 A
0,7 A
0,7 A
ADS 1802
B.T.
200 à 230 V - 50/60 Hz
14 A
18 A
1 A
1 A
H.T.
380 à 480 V - 50/60 Hz
9 A
13A
0,7 A
0,7 A
(1) On ADS 602 (with variator) breaker rate on 10A.