G 01078 - Edition 07 - December 06
3/6
C 90
Use of the serial link (M4 monitoring)
List of controls
Order
Parameter
Response
Action
DLI
XXX
OK ou ERR2
Program the automatic transmission interval in min:1 to 255 sec.
DLR
None
Start up the Data Logger.
ECH
ON or OFF
OK or ERR2
Return all characters received on the serial link in the case of ECHON
HDR
<val> see folio 6/6
OK or ERR3
Change the response heading character.
Input its ASCII value.
IDN
None
V X.X
Send back the software version number.
SET
XXYZZZZ
OK or ERR2
Parameter setting: (example:
# 0 0 0 S E T X X Y Z Z Z Z
)
XX = 00 : ADP power
01 : Aux. temperature
02 : Analog input
03 :
Nitrogen
04 : E1 logic input
05 : E2 logic input
06* : Purge flow prolonged
07 : ADP Maintenance
08 : ADP LP bearings
09 : Load Lock maintenance
10 : Exhaust Maintenance
11 : ROOTS Maintenance
12 : Reserved
13 : Reserved/MFS alert
14 : Date setting : Day
15 : Date setting : Month
16 : Date setting : Year
17 : Time setting : Hour
18 : Time setting : Minute
19 :
Free
20 : Set. (0)/ADP T° alarm (1)
21 :
Reserved
22 : E3 logic input
Y = 0 : Alert
(concerns parameters
1 : Alarm
XX=00/01/02/03/04/05/22)
ZZZZ =
Threshold value (see page 5/6)
SEP
<val> see folio 5/6
OK or ERR3
Used to change the separator character between 2 consecutive data
items in a response. Input its ASCII value.
SYS
ON or OFF
OK or ERR2
Start up the pumping unit with SYSON and stop with SYSOFF.
TPS
XX
Maintenance time
XX = 00 : Running time
01 : ADP Maintenance
02 : ADP LP bearing
03 : Load Lock maintenance
04 : Exhaust Maintenance
05 : ROOTS Maintenance
06 : Free
ROO
ON or OFF
OK or ERR2
Start up the ROOTS with the command ROOON and stop with
ROOOFF.
PUR
ON or OFF
OK or ERR2
Start up the purge with PURON and stop with PUROFF.
* When XX=06 then
Y=0 : purge flow prolonged for 10 minutes, setting values from 0 to 120 min.
Y=1 : standby purge for 10 seconds, setting values from 0 to 9999s.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two