G 02273 - Edition 04 - September 14
1/3
D 10
First level of maintenance
Maintenance time
monitoring
Depending on the maintenance parameter programmed, when the
operating time before maintenance has expired, the screen display
alternates between operating parameters and the corresponding
warning message.
The red indicator light is lit on front panel.
This information can be read at any time using the MAINTENANCE
menu.
Perform maintenance on the relevant parts.
Maintenance
frequency
Maintenance times can change according to processes and equipment
used. Contact the pump manufacturer for more information.
Oil change
There is no oil change required on site. This is done during the pump
overhaul by the manufacturer’s trained personnel.
When a problem occurs, the user is informed by:
the relevant fault indicator light which is located on the front panel:
yellow lit in case of warning, red lit in case of fault (followed by a
pump stop or not, depending on setting parameters, see
C 42
),
the audible buzzer warning (if preselected),
the fault contacts on the J14 remote connector at the rear of the
monitoring system which close,-
the stopping of the pumping although the STOP command has not
been given,
the display of fault message on the display of the hand-held remote
control,
a message via the RS 232 or RS 485 link on a micro-computer.
Study the safety instructions related to maintenance (See
D 00
).
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two