G 01059 - Edition 07 - Januar
y 10
1/4
B 70
J 14 remote control plug connection
In local mode, the pump can run only if the cover plug (packaged with the
pump) is fitted on “J 14” connector.
Location of J 14
remote control
connector
The remote control connection is done via the
“J 14
“ connector,
located on the electrical rear panel of pump.
The remote control function allows:
Remote control of pumping functions “START/STOP/PURGE”,
“Roots“ and “Inlet valve“.
Remote monitoring of parameters through auxiliary dry contacts
(50V - 1A). These contacts can be used to control automatic
functions.
J 14
Male Sub D - 50 pin
connector
Dry contact outputs are rated 50 V - 1 A max.
Overvoltage and overcurrent can cause internal electrical damages.
Monitor external contacts voltage below 50 V and current below 1 A.
When units containing control circuits are equipped with dry contact
outputs, it is the responsability of the customer to use these outputs in
compliance with extra low voltage installation and security standards: it
concerns J1, J6 and J 14 connectors.
Study the preliminary precautions (see
B 00
).
Make sure that the main switch is off to avoid any short circuit during
remote control plug disconnection/connection.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two