2/2
E 012
Draining of the water cooling circuit
G 01159 - Edition 05 - December 06
Draining the
water volume
Turn on the pump (main switch to 1).
In the monitoring DEFINITION menu, set
THERM. ADP OPTION
to
INVALID.
Press
START
. Inject compressed air into the
IN
inlet and let the pump
run until all the water has been completely evacuated (or check
whether there is any water in the flowmeter).
Press
STOP
and confirm by pressing
ENTER
.
Enter monitoring again to set
THERM. ADP OPTION
to VALID.
When the water is filled out, quickly stop the pump and disconnect
pipes (
A
) and (
B
).
This operation must be performed quickly, because without any water
cooling the pump will overheat after a few minutes and will be in
warning/hazard mode.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two