G 01054 - Edition 08 - Januar
y 10
1/2
B 40
Inert gas purge connection
(N
2
connection)
Nitrogen
characteristics
For maximum performances,
a filtered dry nitrogen supply
with the
following characteristics is required:
Condensation point < 22°C
Dust < 1µ.m
Oil < 0.1 ppm
H
2
O concentration < 10 ppb
O
2
concentration < 5 ppb
Pressure between 3 and 7 absolute bar (43 /100 psi)
Flowrate per pump (see
A 40
)
Connection
4 mm
6 mm
1/8 NPT
1/8 NPT
Pipe 6.35 mm OD
1/4 inch tubing
(supplied by
the customer)
Separately
packaged
Facilities on the top
Facilities
at the rear
A sudden inrush of N
2
into the pressurization circuit may result in damage
to internal parts. Close the pressure regulator by turning the knob counter-
clockwise before pressurizing the circuit.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two