Conditioning the pump for shipping
Pumps that are to be shipped must first be pressurized with dry
nitrogen (see
E 11
).
Preparing the pump
for expedition
Install the connections provided with the pump when it was first
delivered. Contact the service center for components if necessary.
Inlet
You must have a supply of dry nitrogen to pressurize the pump.
See
B 40
for the dry nitrogen’s characteristics.
Exhaust
Accessory
DN 50 ISO-KF
(ADP)
DN 100 ISO-K
(ADS)
Inlet blank-off flange
106314
090467
Equipped centering ring
087168
068349
Quick-connect clamp
303003
–
Double-claw clamps (set of 4)
–
303056
1/8 gas connector
082981
082981
Accessory
DN 40 IS0-KF
Blank-off flange
068197
Equipped centering ring
068230
Releasable clamp
303002
Refer to the preliminary precautions before intervene on the product
(see
E 14
).
Pressurizing the pump
Block the inlet port with the flanges supplied.
Connect the nitrogen to the gas connector.
– Pressurize the pump with dry nitrogen to an ultimate pressure of
1.5 bar.
– When the nitrogen flows out of the exhaust, seal it shut with the
accessories provided.
– Disconnect the nitrogen supply.
Fill the safety questionnaire sheet