2/2
B 30
Connection to the cooling circuit
G 01053 - Edition 13 - September 14
Remove the front cover.
The water flow will be adjusted according to the pump model
(see
A40
) with the needle tap located on the flowmeter panel.
Water flowrate
adjustment
Flowmeter panel
Needle tap
Water flowrate
Install the two quick-connect connectors, packaged separately, on the
self-sealing quick-connect connectors.
Connect the water pipes to the quick-connect connectors.
- Water inlet marked ”
IN
” (1/4 inch NPT quick female connector)
- Water outlet marked ”
OUT
” (1/4 inch NPT quick male connector)
ALCATEL supply
Adaptator NPT/BSPP
Facilities on the top
Facilities
at the rear
1/4 NPT
1/4 BSPP
1/4 NPT
1/4 BSPP
Water cooling quick
connection
Separately packaged
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two