J 14 remote control plug connection
All dry contacts are normally opened.
Control contacts
(inputs)(cont’)
Contact Function
S1
(18-19)
Local mode or
Remote control
mode
Contact closed, local mode.
Contact open, remote control mode.
S2
(20-21)
Pump Start/Stop
Contact closed, ADP starts.
Contact open, ADP stops.
This function is only valid when remote control mode is validated with
contact S1 open.
S3
(22-23)
Roots Start/Stop
Contact closed, Roots starts.
Contact open, Roots stops.
This function is only valid when remote control mode is validated, with
contact S1 open, and when “
ROOT.CMD
“ selection is validated in
DEFINITION menu of hand-held remote control.
S4
(24-25)
N
2
purge
Start/Stop
Contact closed, N
2
purge Starts.
Contact open, N
2
purge Stops.
This function is only valid when remote control mode is validated, with
contact S1 open, and when “
PURGE.CMD
“ selection is validated in
DEFINITION menu of hand-held remote control.
S5
(26-27)
N
2
stand purge
valve Start/Stop
Contact closed, N
2
stand purge starts after a time set (time
programmed by RS232, see
C 90
).
Contact open, N
2
stand purge stops.
This function is only valid when remote control mode is validated, with
contact S1 open, and when “
N2 STANDBY
“selection is validated in
DEFINITION menu of hand-held remote control.
S6
(28-29)
Opening/closing
of inlet valve if S7
is activated
Contact closed, closing of inlet valve.
Contact open, opening of inlet valve.
This function is only valid when:
• the remote control mode is validated
• ”
INLET VALVE OPTION
” selected and validated in the DEFINITION
menu of hand-held remote control.
• Inlet valve control authorization is not validated (S7 contact open).
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two