4/6
C 90
Use of the serial link (M4 monitoring)
G 01078 - Edition 07 - December 06
Order
Parameter
Response
Action
DEF
XXY
FAULT
Display the 10 latest Alerts or Alarms
(Example :
# 0 0 0 D E F X X Y
)
XX = 0 to 09 (00 the newest, 09 the oldest)
Y = 0 alarm, Y = 1 alert
OPT
XXY
Choice of the pressure and the purge flowrate
XX = 20, Y = 0 : standard pressure, Y = 1 : specific pressure
XX = 21, Y = 0 : MFS = 60 slm, Y = 1 : MFS = 120 slm
Dual cooling on ADS 1802 H
XX = 26, Y = 0 : DISABLED, Y = 1 : ENABLED
Tansmission of a Data Logger frame
Example :
#000 00100010000 0990 0 080 017 200 000 000 00 000000000000000000000 0000
0/3 5/15 17/20 22/25 27/29 31/33 35/37 39/42 44/45 47/67 69/72
STA
NONE
0 :
header
character
1-2-3 : address
4 :
separator
character
5 :
ADP
status
6 :
Roots
status
7 :
purge
status
8 :
stand-by
9 :
water
valve
10 :
inlet
valve
11
: authorization valve closed
12
: ADP water valve
13
: Roots water valve
14
: pump LL on/off
15
: monitoring mode control*
Bit 7 : always to 1
Bits 6,5,4,3,2,1 : reserved
Bit 0 = 0 : local mode
Bit 0 = 1 : remote mode
16 :
separator
character
17-20 : pressure (mbar)
21 :
separator
character
22-25 : power (W)
26 :
separator
character
27-29 : ADP temperature (°C)
30 :
separator
characterr
31-33 : reserved (°C)
34 :
separator
character
35-37 : auxiliary temperature (°C)
38
: separator character
39-42 : analog input (mV)
43
: separator character
44-45 : MFS (max. 99 slm)
46
: separator character
47
: pressure fault
48
: aux. temperature fault
49
: consumption fault
50
: E analog fault
51
: E1 logic fault
52
: E2 logic fault
53
: variator fault
54
: water flow fault
55
: main power supply fault
56
: motor temperature fault
57
: valve fault
58
: breaker fault
59
: N
2
purge fault
60
: maintenance fault
61
: ADP high temperature fault
62
: ADP low temperature fault
63
: Roots temperature fault
64
: LL pump fault
65
: E3 logic fault
66 :
free
fault
67 :
free
fault
68 :
separator
character
69-72 : N
2
purging in second
73 :
CR
List of controls
(continued)
Bit 7 = 1
* Monitoring mode control (15) :
reserved
1
Bit 0 = 0 : Loc
Bit 0 = 1 : REM
X
Item 5 to 14 :
= 0 : OFF
= 1 : ON
Item 47 to 65 :
= 0 : OK, no fault
= 1 : alert
= 2 : alarm
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two