G 01049 - Edition 10 - Januar
y 10
3/3
B 10
Unpacking / Storage
WARNING
HEAVY OBJET
Can cause muscle strain or back injury.
Use lifting aids and proper lifting techniques
when removing or replacing.
Handling
Regarding its weight, handling the pump can cause muscle strain or back
injury.
For all equipment handling, use the appropriate handling devices.
Handling the pump to
remove it from the crate
Prepare the pump
Screw in the leveling pads of the frame and lower the pump on the
wheels.
Remove any additional packages from the crate and set aside.
These packages will contain cables, accessories, and so forth.
To prevent humidity from entering the pump during transport, the unit
has been pressurized with dry nitrogen before shipment. The inlet and
exhaust are sealed with blank-off flanges which should not be removed
until the pump is ready to be used.
Hook
Hoisting
ring
Equipment storage
It is highly recommended that a hoist
is used for lifting.
Fit the hoisting rings (delivered in
the crate in separate packaging) in
the appropriate holes.
With a hoist and a sling with three
arms, lift the pump using hoisting
rings.
Pump
Weight (lbs/kg)
ADP 122
535 /
243
ADS 602
833 /
378
ADS 1202
1170 /
531
ADS 1802
1097 /
498
When the pump is new, if it is going to be put into storage, the inlet
and exhaust seals should be left in place because the pump has been
pressurized with dry nitrogen to prevent humidity from entering.
Our equipment can be stored without particular storage precautions
(pump pressurized in nitrogen and sealed) only at an ambient temperature
between - 25°C and + 55°C.
The hoisting rings are delivered with the pump, screw them in the
appropriate holes on the top of the pump. They only must be use to handle
the pump.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two