Pump or Roots does not start
Incident
Cause
Troubleshooting
Only ADP starts
No “Roots“ message on the display
Incorrect M4 configuration
Configure for “ADS“.
Roots command “VALID“
Reset the configuration if separate
Roots command is not necessary.
Configure Roots command “INVALID“.
Press “ROOTS“ key
Roots does not run but “ROOTS“
message on the display
Roots LED off
Check the presence or connection of
RP plug located on electronic inside
cabinet
J12
(see
D 10
).
Other problem
Call customer service center.
Display is showing other messages
see
D 500
for trouble-shooting