Safety instructions
Other located hazardous energies
Water cooling circuit and nitrogen purge are pressurized hazardous energies.
Release pressure before servicing:
for the N
2
purge, disconnect the gas line quick connector and turn off
the pressure regulator by turning the knob counter-clockwise;
for the water cooling circuit, disconnect the input connector and leave
the output connected. Wait 1 minute before disconnecting the output.
Operating conditions may generate temperatures justifying particular
attention on the part of the user (external surfaces > 70°C on exhaust
connections).
Contact may cause burns.
Always use protective gloves and wait 30 mn before servicing.
When the pump is switched off, internal parts (monitoring, frequency
converter) contain capacitors charged with over 60 VDC and remain
energized.
Electrical shock may result in severe injury.
Wait 1 minute after switching off before opening the pump.
Lock out/tag out (LO/TO) water and N
2
supplies.
The end user needs to provide a dedicated source of water and Nitrogen
equipped with a manual valves that can be locked out within 10 feet of the
tool.
HOT SURFACE
Contact may cause burn.
Do not touch or wear protective gear before
servicing.
WARNING
The user and /or OEM are ultimately responsible for operating the
equipment in a safe manner. The manufacturer has no control over the
types of gases exposed to this pump. This is the user and/or the OEM’s
responsibility to follow the necessary safety requirements.
Frequently process gases are toxic, flammable, corrosive, explosive and/or
otherwise reactive.
Toxic gases can cause serious injury or death. Operators and users must
take the appropriate safety recommendations to prevent injury. Consult the
responsible department for instructions and safety information.
Hazard gases through the pump can cause serious injury or death. It’s
mandatory by regulations to connect the pump’s exhaust to a facility
hazardous gas exhaust system which incorporates appropriate filters,
scrubbers, etc. This system must meet all air and water regulations. Check
that pump is correctly connected to the equipment (
B 60
).
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two