2/3
D 400
Pump or Roots does not start
G 03218 - Edition 01 - December 06
Incident
Cause
Troubleshooting
Hand-held remote control display
is showing
Hand-held remote controlis on soft
loading configuration
Set the switch
S1-1
to “
OFF
”
(see
C 10
).
Switch off for 10 s, then switch on
again.
Other problem
Call customer service center.
Hand-held remote control display
is showing
Electronic is in “slave” mode:
without hand-held remote control
Set the switch
S2
to “
0
”.
Switch off for 10 s, thenswitch on
again. (see
D 10
).
Other problem
Call customer service center.
Display is showing
… by pressing on any key …
Pump is used with several
hand-held remote controls
Only one hand-held remote control
can monitor the pump.
Press on
CONTROL
to disable the
monitoring on other remote controls.
Press on
CONTROL
on the requested
remote control to monitor the group.
Wait for the display:
Other problem
Call customer service center.
Message on display
Pump in storage mode
Set the main switch to “
0
” to switch
off the pump.
Switch it on again.
Message on display
In Remote mode:
No cover plug or bad connection
Check the wiring of J14 connector
(see
B 70
).
M4 problem
Call customer service center.
MONITOR M4 D 9111
VX.XX
MONITOR M4 D 9111
VX.XX
WAIT ???
WAIT ???
XXH
LOC.N2.
PUMPING STOPPED
MM/JJ/AA HH:MM
XXH
LOC.N2.
PUMPING STOPPED
MM/JJ/AA HH:MM
XXH
LOC.N2.
COMMAND NOT ALLOWED
JJ/MM/AA HH:MM
XXH
LOC.N2.
COMMAND NOT ALLOWED
JJ/MM/AA HH:MM
* XXH
LOC.N2.
PUMPING STOPPED
MM/JJ/AA HH:MM
* XXH
LOC.N2.
PUMPING STOPPED
MM/JJ/AA HH:MM
STOCK
STOCK
REM
REM
Display is showing a message or an alarm