G 01078 - Edition 07 - December 06
5/6
C 90
Use of the serial link (M4 monitoring)
Modifying the
separator character
Order
Parameter
Response
Action
SEP
<val>
val = decimal value
of the ASCII code
of the separator
character between
1 and 255
OK or ERR3
Change the separator character between two consecutive data items
in the response.
Note:
- The factory configuration is the decimal code «032» of the
«space» character.
- Customer selection is memorized by the monitoring.
An order of the type:
#000SEP044<CR>«044»
is the decimal ASCII
code of the character «,» (comma).
Command
#000STA<CR>
Response
#000,00000,2,0080,2530,0127,1230,0000000000 <CR>
Only commands requiring the separator character are affected by the
SEP order.
Example:
Table for mini/maxi
parameter thresholds
Parameter
Unit
Thresholds
Mini
Maxi
Limit
Mass-flowmeter
slm
0
60
(P version)
120
(H version)
ADP pump temperature sensor
°C
70
130 or
140
(ADS 1802)
Moter power sensor
W
500
5000
E. analogic sensor
mV
0
9950
LI1. logic sensor
mn
0
60
61
DISABLED
LI2. logic sensor
mn
0
60
61
DISABLED
LI3. logic sensor
mn
0
60
61
DISABLED
Auxiliary temperature
(dual cooling on ADS 1802 H)
°C
70
100
Auxiliary temperature ADS 602 P - ADS 602 H
°C
0
170
Auxiliary temperature
°C
70
170
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two